Detailed description
PARAMETERS | VALUES |
Measurement uncertainty CALIB100 | 0.2 % of the final value (for measuring range > 0 … 200 Pa/± 200 Pa) or 0.5 % de la mesure finale (for measuring range ≤ 0 … 200 Pa/± 200 Pa) |
Measurement uncertainty CALIB200 | 0.1 % of the final value (for measuring range > 0 … 200 Pa/± 200 Pa) or 0.2 % of the final value (for measuring range 0 … 200 Pa/± 200 Pa) or 0.3 % of the final value (for measuring range 0 … 100 Pa/± 100 Pa) |
Hysteresis | 0.1 % of the final value |
Overload capacity | 600 kPa (for measuring range > 3 kPa) 200 fois (for measuring range ≤ 3 kPa) |
Zero point thermal drift | ± 0% (cyclic zero point correction) |
Influence of temperature (margin) | CALIB100 : 0.04 % of the final value / K (10 … 40 °C) CALIB200 : 0.03 % of the final value / K (10 … 40 °C) |
Calibration temperature | 22 °C |
Fluids | Air, all non-aggressive gases |
Measurement input / Power supply | 0 … 10 V, 0/4 … 20 mA Measurement accuracy : 0.2 % of the final value 24 VDC / 100 mA |
Display | Alphanumeric display with 2 x 20 characters, backlight |
Operating temperature | 10 … 40 °C |
Storage temperature | -10 … 70 °C |
Weight | About 4.5 kg |
Pressure fittings | Ø6 mm for DN 5 mm pipe |
Certifications | CE |